Algorithm and Functional Scheme of Data Acquisition at Test Bench for Studying Thermal Conductivity at Temperatures below 10 K
Published: 15.07.2013
Authors: Arharov I.A., Koshelev S.S., Sergatskov D., Carcagno R.
Published in issue: #1(13)/2013
DOI: 10.18698/2308-6033-2013-1-601
Category: Engineering | Chapter: Cryogenic Engineering
For studying thermal properties of superconducting niobium, it is necessary to provide temperatures within a range of 1.4 to 10 K and adequate aids for their measuring. An error of temperature measurements must not exceed 1 mK. Types of thermometers used at low temperatures are examined. Zirconium oxynitride is chosen as the basic thermometer material. An algorithm for measuring is proposed which makes it possible to reduce errors of temperature measuring and improve its accuracy.